StatSpin® Express 4 Horizontal Centrifuge Operator’s Manual A Division of IRIS International, Inc.

Pagina 2 - Model Number M510

6 Section 3 Operating Instructions Opening and Closing the Cover An electronic operated cover interlock mechanism prevents operation if the cover

Pagina 3 - Table of Contents

7 2. Locate 1/8 inch hex key supplied and remove center screw of rotor hub. 3. Remove rotor lifting straight up and out of centrifuge. To re-ins

Pagina 4 - How to use this manual

8 Section 4 Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenanc

Pagina 5 - Unpacking and Installation

9 Periodic Maintenance Check When the cycle counter in the instrument reaches 50,000 spins, the Error/Service indicator will continuously illuminate

Pagina 6 - System Overview

10 Troubleshooting Symptom: Solution: Red Error / Service indicator flashing, unit will not spin. Refer to Error Code Table of manual to identify a

Pagina 7 - Error Indicators

11 Service Refer all service to Iris Sample Processing or an Authorized Repair Facility. Instrument fuse(s) are located on the rear panel. Remove

Pagina 8 - Accessories

12 Section 5 Appendix Specifications Product Number SSH4 Model Number M510 Spin Parameters 3 min: 180 seconds @ 5100 rpm / 4000 xg 5 min: 300 sec

Pagina 9

13 References 1. CLSI. “Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline-Third Edition.” CLSI documen

Pagina 10 - Operating Instructions

14 Iris Sample Processing Warranty Iris Sample Processing, a Division of IRIS International, Inc. warrants that the centrifuge, with the exception

Pagina 11 - Unloading

Operator’s Manual StatSpin® Express 4 Horizontal Centrifuge Model Number M510 For in-vitro diagnostic use Product Number SSH4 – StatSpin® Exp

Pagina 12 - Maintenance

i Table of Contents How to use this manual…………………………………………..……….………… ii Section 1 ……………………………………………………………..……….……….. 1 Unpacking and Installatio

Pagina 13 - Periodic Maintenance Check

ii How to use this manual This manual along with information contained on product labels and in package inserts should provide you with all the inf

Pagina 14 - Troubleshooting

1 Section 1 Unpacking and Installation Inspect Packaging The StatSpin® Express 4 and its accessories are delivered in a single carton. If the centr

Pagina 15 - Service

2 Section 2 System Overview Principle and Intended Use For in-vitro diagnostic use to produce the rapid separation of whole blood contained in origin

Pagina 16 - Appendix

3 Locked Locked position for shield Unlocked Unlocked position for shield Operator Controls Start The Start button initiates the selected cycle

Pagina 17

4 Error Codes Table Failure to reach speed in 30 seconds. Confirm no obstruction in bowl area, contact customer service. Latch is not lock